6. To program minimum graph percentage,
the display will indicate 0 0 with the right
digit flashing, the bottom most segment
of the bar graph will flash, and the dis-
play will read “PCTFS (percent full scale).
Press
,
to scroll to a num-
ber 0-9. Press
|
|
to select that num-
ber. The left digit will now begin to flash.
Press
,
. to scroll to a
number 0-9.
|
|
. to select that
number. The 2 digit number entered
represents the percentage of full scale
to be used as the low end of the graph
(0-99%).
7. To program maximum percentage of full
scale the display will indicate 1 0 0 with
the right digit flashing, the upper most
segment of the graph will flash, and the
display will read PCTFS (percent full
scale). Press
,
. to scroll to
a number 0-9. Press
|
|
to select
that number. The left 2 digits will now
begin to flash. Press
,
to
scroll to a number 0-10. Press
|
|
to select that number. The number en-
tered represents the percentage of full
scale to be used as the high end of the
graph. Note: 100 is the highest and only
possible 3 digit number. If the low num-
ber is equal to or larger than the high
number, the unit will flash “REJCT”
and it will exit back into the “GRAPH”
sub-menu.
Note: Custom graph function is not available
on vacuum and compound ranges.
Note: A display message “ULOCK” indicates
that custom units are being used and bar
graph minimum and maximum are set to
the factory defaults
Reset:
Returns the product to the factory default val-
ues.Preserves field calibration. Factory calibra-
tion can be restored in the “RECAL menu.
1. In menu mode
,
until “RESET”
appears on display’s lower line;
|
|
.
Factory defaults pertain to units, timer,
back light, update rate, zero lock.
Display will indicate “DONE” then gauge
will switch to “OFF condition.