20
Maximum display span: OFFSET to FULL SCALE
For display mode, select “Linear display mode”
⇒ Setting item
Set min. pressure P1 as “10 psi”
⇒ Setting item
Set max. pressure P2 as “60 psi”
⇒ Setting item
Set decimal point position of linear display
as “two digit”
⇒ Setting item
Set OFFSET of linear display as “0.00” ton
⇒ Setting item
Set FULL SCALE of linear display as “5.00” ton
⇒ Setting item
Set output zero point as “0.0%F.S.”
(0.00ton) of max. display span
⇒ Setting item
Set output span point as “100.0%F.S.”
(5.00ton) of max. display span
⇒ Setting item
Linear display and Analog
output
Linear Analog
Pressure Display Output
(psi) (ton) (mA)
60 ==> 5.00 ==> 20 Span Point
兰兰 兰
35 ==> 2.50 ==> 12
10 ==> 0.00 ==> 4 Zero Point
ton