18
For display mode, select “Linear display mode” ==>Setting item
Set min. differential pressure P1 as “20 in.H
2
O” ==>Setting item
Set max. differential pressure P2 as “120 in.H
2
O” ==>Setting item
Set decimal point position of linear display
as “one digit” ==>Setting item
Set OFFSET of linear display as “0.0m” ==>Setting item
Set FULL SCALE of linear display as “50.0m” ==>Setting item
Set output zero point as “0.0%F.S.”
(0.0m) of max. display span* ==>Setting item
Set output span point as “100.0%F.S.”
(50.0m) of max. display span* ==>Setting item
*Maximum display span: OFFSET to FULL SCALE
䡵 Linear display and analog output
Ex. Differential pressure 70 in.H
2
O
• Linear display: 25.0(m)
• Analog output: 12(mA)
m
(1)
Differential Linear Analog
Pressure Display Output
(in.H
2
O) (m) (mA)
120 ==> 50.0 ==> 20 Span Point
兰兰 兰
70 ==> 25.0 ==> 12
20 ==> 0.0 ==> 4 Zero Point
(1) Put the arbitrary unit sticker.
See Section 14.4 for full menu.