Set the output span point as “100.0%F.S.”
(120.0 arbitrary unit of measure) of
max. display span* ==> Setting item
Set time factor selection
(seconds, minutes, hours) ==> Setting item
Select standard “I” as flow rate volume factor ==> Setting item
21
Select “ti” (automatic) as display
switching method ==> Setting item
22
Set automatic display switching time
as “5 seconds” ==> Setting item
23
*Maximum display span: 0 to Maximum momentary flow rate
84.9
270225
Arbitrary unit of square root display and analog
Momentary flow rate display (standard)
Outbreak differential pressure: 16IW Integrated volume display (arbitrary unit)
• Second flow quantity: 84.9 Integrated volume 270225
(Next page reference)
• Analog output: 15.3 mA
It It is alternation
indication for every
five seconds
(Automatically)
Differential Pressure Momentary Flow Rate Unit Analog Output
(
IW) (mA)
Span Point
32 120 20
|||
16 84.9 15.3
|||
Zero Point
004